L-0.2, r. 1 - Regulation respecting the application of the Act respecting medical laboratories and organ and tissue conservation

Full text
157. The attenuation of x-rays by the surface of the examination table must not be more than the equivalent of 1 mm of aluminium measured at 100 kV (peak voltage) where the tube is above the table.
R.R.Q., 1981, c. P-35, r. 1, s. 157.