L-0.2, r. 1 - Regulation respecting the application of the Act respecting medical laboratories and organ and tissue conservation

Full text
124. In a medical biology laboratory:
(a)  readily accessible pressure showers must be made available to the staff when the latter is exposed to toxic or corrosive substances;
(b)  fire blankets must be available at all times in well identified and readily accessible places. Masks and gloves for toxic products must be available for members of the staff when their safety so requires;
(c)  the private and isolated compartment prescribed in paragraph d of section 122 must contain a bed or a stretcher as well as a call system within easy reach;
(d)  safe storage rooms must be provided for the storing of chemical or biological products in order to avoid the particular hazards caused by their volatility or their explosive nature.
R.R.Q., 1981, c. P-35, r. 1, s. 124.